Special proiectat pentru o utilizare facila, Echo stabileste noi standarde in simplitate, astfel minimizand timpul de instruire al personalului si lucrul efectiv. Echo are o viteza intre 120-240 teste de biochimie pe ora fiind destinat laboratoarelor mici si medii cu 64 probe si 27 de reactivi la bord. Metode de citire sunt end point, kinetic si fixed time, cu factor sau standard, mono- sau bi-chromatic, curbe de calibrare liniar si non-ineare cu 8 standarde, cat si teste turbidimetrice. Echo are precizie inalta si repetabilitate datorita utilizarii unor sisteme electronice si mecanice complexe ce necesita un minim de intretinere.
CARACTERISTICI TEHNICE Pozitii reactivi: 18 sau 27 pozitii de reactivi pe placa rece Viteza: 200 - 270 teste / ora (12 sec pentru fiecare ciclu) Incarcarea cu probe: 64 probe, standarde si controale Metode: 64 metode in memorie. Teste cu 2 reactivi. End Point la o lungime de unda sau doua, citire diferentiala, blank de proba, Kinetic, Fixed Time, metode cu unu, doua sau trei reactivi Calibrare: Lineara si non lineara, cu pana la 8 standarde - factori Sistemul optic:
- Aducere la zero automata pentru toate lungimile de unda
- Acuratete: +/- 1 % de la 0 la 2.5 O.D.
- Liniaritate: < 0.5 %.
- Doua celule de citire cu volum 35 microlitri.
- Temperatura controlata de elemente Peltier (+/- 0,1 oC).
- Aspiratie prin pompa peristaltica programabila
Modul dilutor: Seringa de 1000 microlitri cu rezolutie 0.3 microlitri. Aducere automata la zero Placa de reactie: 190 microcuve la temperatura de incubare 37oC, Programabil +/- 0,1 oC Interfata utilizator: Tastatura, Monitor color 15”, Imprimanta termica incorporata cu 24 coloane, optional imprimanta externa. Posibilitate de comunicare bidirectionala cu PC extern (Optional se livreaza si softul de comunicare). Inregistrarea completa a datelor pacientului. Program de control al calitatii, Calcul statistic. Rezultate afisate si printate, Soft de autodiagnosticare. Interfata RS232.
SIZE / WEIGHT Height: 35 cm (13.8 in)
Depth: 67 cm (26.4 in)
Width: 70 cm (27.6 in)
Weight: 45 kg (99.2 lbs)
POWER SUPPLY 240 or 110 Vac, 50 / 60 Hrz, single phase with ground
Fuse compartment
Fuses: 3.15 Amp @ 230 Vac, 6.3 Amp @ 115 Vac
Power consumption: less than 300 VA (external PC excluded)
Ground resistance: less than 0.1 Ohm
Leakage current: less than 2.5 mA
HYDRAULIC
1 aspiration needle
SYSTEM
1 self-priming peristaltic pump
Pinch valve
Containers: Water, 500 ml and waste 1000 ml
REAGENTS TRAY
3 removable racks
27 bottles of 25 ml (675 ml total),
or 18 bottles of 40 ml (720 ml total)
SAMPLES TRAY
Standard removable tray:
1 blank position,
63 numbered positions for cups,
16 positions for standards and controls
REACTION
96 replaceable cuvettes (four segments of 24 cuvettes)
CUVETTES ROTOR
500 μl reaction volume
100W heating resistance, temperature sensor, safety thermostat
OPTICAL GROUP
1 halogen lamp, 6 V / 10 W with extended UV emission
2 focusing lenses, optical glass
Filter disk: 8 positions provided with interference
filters of 340, 405, 492, 505, 546, 578, 630, 700 nm wavelengths,
1 free position and 1 solid position for dark reading
1 quartz flow cell, 10 mm optical path
±2 nm on peak wavelength, band pass of ±10 nm
PHOTOAMPLIFIER
Photoelectric detector
Response range, 340 nm to 900 nm
Photometric range, 0 to 2.5 Abs ±0.003 Abs
Linearity, ±0.5% full scale
Precision: 1 CV% (0.050 to 1.500 Abs)
Stability: daily reader offset, less than 1% drift per day
CONTROL Real-time multitasking microprocessor based control
Easy access to the electronics
ECHO XPC External computer and Windows software COMPONENTS 17’’ monitor, 1280x1024 resolution
External PS/2 keyboard and mouse
External printer ECHO PC Internal PC COMPONENTS Internal thermal printer, 24 columns
17’’ monitor, 1280x1024 resolution
External PS/2 keyboard and mouse
(Optional) External MS DOS compatible printer
PIPETTING
Volume: sample, 3 - 600 μl; reagent, 3 - 600 μl
Precision: 1.5 CV% at 10 μl; 1 CV% at 250 μl
Mixing by sample needle upon dispensation of R2 and R3
REACTION Reaction volume, 400 - 600 μl SAMPLE Automatic dilution of standards and samples TEMPERATURE Reading flow cell (25 - 42 °C ±0.2 ºC) CONTROL Reagent refrigeration, circa 15 °C below room temperature
Reagent preheating reel, 0 to 100% heating setting
Reaction cells, heating unit can be set from
room temperature up to 42 °C ±0.2 ºC (108 °F ±0.5 ºF)
(with secondary thermostatic safety control)
TYPES OF TESTS Endpoint, Bichromatic endpoint, Differential endpoint, Differential endpoint sample blank, Fixed Time, Kinetic TEST RUNS Batch, urgent MEASUREMENT 100 kinetic tests/hour, 270 endpoint tests/hour RATES Instrument cycle: 12 seconds
Typical precision: 2.0 CV% max for endpoint
and 3.0 CV% max for kinetic
CALIBRATION Reagent blank subtraction; 1 to 8 standards per test method Linear: Factor, linear, linear regression (standard’s repetitions) Non linear (2 interpolation types): cubic-spline, poly-linear
16 standard / control positions on the sample plate
PRINTING Per sample or per test
Full worklist run report TEMPERATURE/ Temperature range, 16 °C to 30 °C (61 °F to 80 °F) HUMIDITY Humidity, 10% to 80% relative with no condensation STORAGE 0°C ÷ +50% at maximum 85% of relative humidity NOISE LEVEL Average 62 dB, peak 64 dB POWER Standard VDE removable power cord HOST/ LIS Standard RS232, external 9-pole male
(test methods, samples, work list, results) TEST METHODS 64 methods stored in memory
Single, double and triple reagent methods QUALITY Three-level controls per test, one month monitoring CONTROL Levey-Jennings graphics